Fluke 5520A High Performance Multi-Product Calibrators

34.500,00 

  • Incl. new Calibration
  • Handheld and bench meters (analog & digital) up to 6 1/2 digits
  • Thermocouple and RTD thermometers
  • Process calibrators
  • Data loggers
  • Strip and chart recorders
  • Watt meters
  • Power harmonics analyzers
  • Current clamps and clamp meters
  • Analog or digital handheld and
    bench oscilloscopes (with options)
  • Panel meters
  • Graphical multimeters
  • Power meters, disturbance analyzers, power quality monitors,
    recorders and other power quality related equipment (with option)
Artikelnummer: 6057 Kategorien: ,

Beschreibung

The 5520A Multi-Product Calibrator sources direct voltage and current, alternating voltage and current with multiple waveforms and harmonics, two simultaneous voltage outputs or voltage and current to simulate dc and ac power with phase control, resistance, capacitance, thermocouples and RTDs. The 5520A can also measure thermocouple temperature, humidity (with probe) and pressure using one of 28 Fluke Calibration 700 Series pressure modules. Three options add the capability to calibrate oscilloscopes up to 1.1 GHz. And the 5520A-PQ Power Quality Option enables the 5520A to calibrate power quality instrumentation to the standards of the IEC and other regulatory agencies.

Inklusive neuer Kalibrierung von Fluke

Wide workload coverage

Because it can calibrate so much of your workload, the 5520A makes your calibration investment go farther. Instead of several instruments to purchase, maintain and transport, the 5520A combines all the capability in one compact, portable and affordable tool.

Easy to use

Though it can do a lot, the 5520A is simple to use. Its front panel is logically arranged and clearly labeled. And its bright LCD displays provide all the information you need at a glance. The 5520A is so intuitively easy to use, you’ll seldom need to consult the manual. For almost every operation, your hand moves naturally from left to right. Values are entered on the calculator-style keyboard. Output values can be stepped up or down easily in even decade multiples (1-2-5 sequence in scope mode) using the multiply and divide keys. You can even view the uncertainty for any value at the press of a key.

Versatile and flexible

The basic 5520A can tackle a tremendous amount of work. And as your workload changes, your 5520A can adapt. Three options are available to address the calibration requirements of oscilloscopes. If your workload is made up mostly of oscilloscopes up to 300 MHz bandwidth, the 5500A-SC300 option is an affordable solution. If your workload extends to 600 MHz, the 5500A-SC600 adds complete, powerful oscilloscope calibration capability to your 5520A. The 5520A-SC1100 option enables the 5520A to calibrate all of your oscilloscopes up to 1.1 GHz with rise times to 300 ps – even 125 ps with the 5800A/TDP tunnel diode pulser option. Any of these options can be fitted at time of purchase, or added later as your needs change.

If three-phase power measurement tools are part of your workload, the 5520A can greatly simplify the task. Its phase locking capability makes it possible to „lock“ three 5520As together to simulate three phase power up to nearly 21 kW with multiple waveforms, harmonics and precise phase control. Using this capability, you can also „sum“ the current outputs of multiple 5520As to source 100A or more to directly verify current clamps or power transformers.

Portable and rugged

The 5520A isn’t limited to the controlled environment of the calibration lab. It’s made to go where the work is. It’s compact so it’s easy to take it where the work is. Its rugged construction means you don’t have to treat it like a delicate lab instrument. And it’s self-contained. No additional test pad is required.

A complete solution

Every 5520A is shipped with a certificate of calibration with results data, traceable to nationally recognized standards. When it comes time to recalibrate your instrument, Fluke service facilities are available worldwide to ensure the work is done right and quickly.

Compliance with quality standards made easy

With quality management standards like ISO 9000, there is a lot more to calibration than just making measurements. You also have documentation, control and reporting requirements to meet.

Our optional Microsoft Windows®-based 5500/CAL simplifies the documentation of your procedures, adequacy and traceability as required by ISO 9000 and other similar quality standards. It also collects and reports calibration results information and helps you to consistently, quickly and efficiently calibrate a wide variety of instruments. With it your entire calibration process — from creating and executing procedures through results data collection and reporting — can be automated. Thousands of procedures — and the list is growing — are included at no extra charge. They can be run as is or used as the basis to create your own new procedures.

5500/CAL is part of a suite of applications called MET/CAL® Plus. 5500/CAL is designed around RS-232 instrument control and operates with the 5500A, 5520A, and 5800A calibrators. If you need IEEE instrument control or want to automate the calibration of additional instruments, you can step up to MET/CAL® . And you can add test and measurement asset management capabilities with MET/TRACK.

All three applications are driven by a powerful, industry standard SQL data base server, SQL Anywhere® from Sybase, providing secure, dependable storage for your calibration information. And each application is completely modular so you can set up a single application on a personal computer or take advantage of its client/server design over your network.

 

  • Handheld and bench meters (analog and digital) up to 6 1/2 digits
  • Thermocouple and RTD thermometers
  • Process calibrators
  • Data loggers
  • Strip and chart recorders
  • Watt meters
  • Power harmonics analyzers
  • Current clamps and clamp meters
  • Analog or digital handheld and bench oscilloscopes (with options)
  • Panel meters
  • Graphical multimeters
  • Power meters, disturbance analyzers, power quality monitors, recorders and other power quality related equipment (with option)

Zusätzliche Informationen

Hersteller

Kategorien

Fluke 5520A Datenblatt

 

5520A High Performance Multi-Product Calibrators Produktdatenblatt

 

Technical Specifications
Voltage DC
Range:  0 to ±1020V
Best 1 Year Specification:  ±12 ppm of setting
Current DC
Range:  0 to ±20.5A
Best 1 Year Specification:  ±100 ppm of setting
Resistance
Range:  0 to 1100 MΩ
Best 1 Year Specification:  ±28 ppm of setting
Voltage AC
Range:  1 mV to 1020V
 10 Hz to 500 kHz, Sine
Best 1 Year Specification:  ±120 ppm of setting
Current AC
Range:  29 mA to 20.5A
 10 Hz to 30 kHz, Sine
Best 1 Year Specification:  ±0.06% of setting
Capacitance
Range:  0.19 nF to 110 mF
Best 1 Year Specification:  ±0.25% of setting
Thermocouple Source and Measure
Range:  -250°C to +2316°C (11 types)
Best 1 Year Specification:  ± 0.14°C of setting
RTD Source
Range:  -200°C to +630°C
 (8 types)
Best 1 Year Specification:  ±0.03°C of setting
DC Power
Range:  10.9 mW to 20.5 kW
Best 1 Year Specification:  ±0.023% of setting
AC Power
Range:  10.9 mW to 20.5 kW
 45-65 Hz, PF=1
Best 1 Year Specification:  ±0.08% of setting
Phase
Range:  0 to ± 179.99°
 10 Hz to 30 kHz
Best 1 Year Specification:  ±0.07° of setting
Frequency
Range:  0.01 Hz to 2.0 MHz
Best 1 Year Specification:  ±25 ppm of setting

 

General Specifications
Warm up Time
 Twice the time since last warmed up, to a maximum of 30 minutes
Settling Time
 Less than 5 seconds for all functions and ranges except as noted
Standard Interfaces
 IEEE-488 (GPIB), RS-232
Temperature Performance
Operating:  0°C to 50°C
Calibration (tcal):  15°C to 35°C
Storage:  -20°C to 70°C
Temperature Coefficient
 Temperature coefficient for temperatures outside tcal ± 5°C is 10% of the 90-day specification (or 1-year, as applicable) per °C
Relative Humidity
Operating:  <80% to 30°C, <70% to 40°C, <40% to 50°C
Storage:  <95%, non-condensing
Altitude
Operating:  3,050m (10,000 ft) maximum
Non-operating:  12,200m (40,000 ft) maximum
Safety
 12,200m (40,000 ft) maximum
Analog Low Isolation
 20V
EMC
 Designed to comply with IEC 61326-1/1997
Line Power
Line Voltage (selectable):  100V, 120V, 220V, 240V
Line Frequency:  47 to 63 Hz
Line Voltage Variation:  + 10% about line voltage setting
Power Consumption
 600 VA
Dimensions
Height:  17.8 cm (7 in), standard rack increment, plus 1.5 cm (0.6 in) for feet on bottom of unit
Width:  43.2 cm (17 in), standard rack width
Depth:  47.3 cm (18.6 in) overall
Weight (without options)
 22 kg (49 lbs.)
Absolute Uncertainty Definition
 5500A and 5520A uncertainty specifications include stability, temperature coefficient, linearity, line and load regulation, and the traceability of the external standards used for calibration. You do not need to add anything to determine the total uncertainty of your calibrator for the temperature range indicated.
Specification Confidence Interval