IET Labs DCCS Series DC Current Shunt

•Resistance material: 100 μΩ – 100 mΩ MANGANIN® sheet, 1 Ω -100 kΩ ZERANIN® wire
•Calibration temperature: 23 °C ± 3 K (< 0.5 W load)
•Surface temperature: max. 85 °C
•Thermal resistance: 11 K/W
•Test voltage: 2900 V DC (resistance element-housing)
•Nominal insulation voltage: 650 V DC (insulated mounting required)
•Insulation resistance: > 100 MΩ
•Specifications: according IEC 60477
•Size: 1.5 x 3.8 x 1.6 in (2.3 in with terminals) Weight: 0.55 lbs

Kategorien: ,

Beschreibung

The IET Labs DCCS Series Current Shunts are calibration grade devices used for precise dc current measurements. They extend high accuracy measurements beyond the available limits of self-contained dc ammeters and the DCCS Current Shunts can be used as standard reference resistors. The DCCS Current Shunts are low cost 4 terminal calibration resistors that combine high accuracy , class 0.02, long term stability and permanence of calibration in a compact unit. Constructed using carefully selected low temperature coefficient Manganin® or Zeranin® wire, depending on the value, and mounted to ensure mechanical stability, these current shunts will provide a cost-effective addition to any laboratory or workshop.

Main Areas of Application For The DCCS Current Shunt:
•For tests of measuring bridges or electrical temperature measuring equipment
•For a laboratory setup of a Wheatstone bridge
•As shunt resistor for current measurement
•As part of standard equipment in research laboratories

Models: DCCS-0.0001, DCCS-0.001, DCCS-0.01, DCCS-0.1, DCCS-1, DCCS-10, DCCS-100, DCCS-1000.