•Nominal Values: R1: 95.3 kΩ ; R2: 5.97 kΩ , R3 374 Ω , 24.9 Ω
•Accuracy: 0.1% of nominal value
•Stability: Better than 25 ppm per year
•Temperature Coefficient 2°C ppm
•Calibration Accuracy: These resistors are compared with a precision of better than 40 ppm at DC. Measurement uncertainty of the nominal Q values are within 100 ppm at 1 kHz. All other R and Q values at other frequencies are mathematically calculated. All calibrations are made at 23° ± 2 C°.

Artikelnummer: 12702 Kategorien: ,

Beschreibung

The IET Labs 7000-09 is a self-contained calibration kit used to calibrate the 7400, 7600 and 7600 Plus Precision LCR Meter. Includes a 1689-9604 BNC to BNC cable set (1 meter in length) for connection to the instrument. Consists of four resistors, a precision open and precision short. A calibration chart traceable to an SI is provided with R & Q values for each resistor at various frequencies.

Features:
•4 Resistors, open and short
•1 meter BNC Cable 1689-9602
•Rugged case
•Calibration Certificate
•Z-Foil Bulk Metal® Resistors